Abstract
Triphenylamine (TPA)-functionalized magnetic microspheres (Fe(3)O(4)/SiO(2)/TPA) were prepared and applied as solid phase extraction (SPE) adsorbents for the analysis of polycyclic aromatic hydrocarbons (PAHs) in environmental samples in combination with high-performance liquid chromatography (HPLC). The magnetic solid-phase extraction (MSPE) conditions affecting the extraction efficiency were optimized, including elution solvent, standing time, amount of sorbent, and salt concentration. Due to the strong π-π conjugate effect between the benzene rings of TPA and PAHs, high extraction efficiency was achieved with spiked recoveries of 80.21-108.33% and relative standard deviations (RSD) of less than 10%. Good linearities (R(2) > 0.997) for all calibration curves were obtained with low limits of detection (LOD) of 0.25, 0.5, 0.5, 3.75, 0.2 and 0.04 ng L(-1) for anthracene, fluoranthene, pyrene, chrysene, benzo[b]fluoranthene and benzo[k]fluoranthene, respectively. The achieved results indicate the applicability of Fe(3)O(4)/SiO(2)/TPA as MSPE adsorbents.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.