Abstract

When it comes to total hip replacements, choosing the suitable material combination is of clinical relevance. The present review article examines the technical differences in wear and corrosion of the relevant material combinations of ceramics, metals, ceramized metals and various types of polyethylene. The material characteristics, which were often tested under standardized conditions in the laboratory, are compared with clinical results on the basis of evidence-graded clinical studies and on the basis of register studies. This article thus represents an up-to-date snapshot of the expectations and actual clinical outcomes of the present choice of material combinations. It shows that some tendencies from the field of materials research, e.g. with regard to cross-linked polyethylene, coincide with observations from practical clinical experience, while for other materials, aproven technical superiority has not yet been confirmed as an evident advantage in clinical practice.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.