Abstract

PurposeStereotactic ablative radiation therapy (SABR) delivered in a single fraction (SF) can be considered to have higher uncertainty given that the error probability is concentrated in a single session. This study aims to report the variation in technology and technique used and its effect on intrafraction motion based on a 10 years of experience in SF SABR. Methods and MaterialsRecords of patients receiving SF SABR delivered at our instruction between 2010 and 2019 were included. Treatment parameters were extracted from the patient management database by using an in-house script. Treatment time was defined as the time difference between the first image acquisition to the last beam off of a single session. The intrafraction variation was measured from the 3-dimensional couch displacement measured after the first cone beam computed tomography (CBCT) acquired during a treatment. ResultsThe number of SF SABR increased continuously from 2010 to 2019 and were mainly lung treatments. Treatment time was minimized by using volumetric modulated arc therapy, flattening filter-free dose rate, and coplanar field (24 ± 9 min). Treatment time increased as the number of CBCTs per session increased. The most common scenario involved both 2 and 3 CBCTs per session. On the average, a CBCT acquisition added 6 minutes to the treatment time. All treatments considered, the average intrafraction variation was 1.7 ± 1.6 mm. ConclusionsSF SABR usage increased with time in our institution. The intrafraction motion was acceptable and therefore a single fraction is an efficacious treatment option when considering SABR.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call