Abstract

We describe a method to calculate the relative biological effectiveness in mixed radiation fields of therapeutic ion beams based on the modified microdosimetric kinetic model (modified MKM). In addition, we show the procedure for integrating the modified MKM into a treatment planning system for a scanned carbon beam. With this procedure, the model is fully integrated into our research version of the treatment planning system. To account for the change in radiosensitivity of a cell line, we measured one of the three MKM parameters from a single survival curve of the current cells and used the parameter in biological optimization. Irradiation of human salivary gland tumor cells was performed with a scanned carbon beam in the Heavy Ion Medical Accelerator in Chiba (HIMAC), and we then compared the measured depth-survival curve with the modified MKM predicted survival curve. Good agreement between the two curves proves that the proposed method is a candidate for calculating the biological effects in treatment planning for ion irradiation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.