This paper reports a novel approach to screening multi-class pesticides by ion mobility time-of-flight mass spectrometry detection. Nitrogen was selected as mobility gas. After optimization of the different ion mobility parameters, determination of matrix effect on the drift times was conducted using different matrix extracts. The results showed that drift time values are not influenced by the matrix and also are independent of the concentration within the working range for 100 pesticides tested, making drift time a powerful additional identification tool. Based on statistics, 2% variation criteria provides a good fit for all the pesticides targeted, and could be considered as a maximum acceptable criteria associated with the drift time parameter for identification purpose. This 2% value is in agreement with already reported criteria, for instance, for GC or LC retention time in European documents. Finally, the well-known feature of mobility to separate complex mixtures was also tested to obtain purified extracted mass spectra of pesticides present in fruit extract.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.