Abstract

BackgroundPlugging of the Eustachian tube (ET) is an effective method for treating patulous Eustachian tube (PET). However, no material suitable for plugging is commercially available. A plugging material was made using an angiocatheter and a ventilating tube (VT). MethodsAn 18-gauge angiocatheter was cut 25 mm from the tip, and the cut end was occluded and widened by melting using a candle. The angiocatheter was inserted into the hole of a Paparella type 1 VT, which was moved to the cut end of the catheter. The VT acted as a stopper at tympanic orifice of ET to prevent the angiocatheter from descending into nasopharynx. Two women with PET received ET-plugging surgery using this plugging material via trans-tympanic approach. ResultsAll symptoms of PET disappeared after surgery. There have been no postoperative complications, and the catheter has functioned well without extrusion. ConclusionA plug created from an angiocatheter and a VT is easy to make and is effective for the treatment of PET.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.