Abstract

Gastric ablation is emerging as a potential therapy for electrical dysrhythmias associated with gastric disorders. Thermal conduction properties of gastric tissue during ablation have not yet been defined, but are necessary for optimizing the technique and translating ablation to clinical therapy. We developed custom needle-based transmural temperature probes to quantify the temperature of gastric tissue during ablation. These probes were applied in vivo in pigs (n=5), during gastric ablation (70 °C, 10 s duration), at distances of 2.5 - 20 mm from the ablation catheter tip. Thermal response of the tissue was non-linear; the maximum temperature increase from baseline (33.3 ± 1.0 °C) was observed at the closest temperature probe to the catheter tip (2.5 mm, 14.9 °C), and temperature change decreased with distance from the catheter tip. Probes positioned between 5 -20 mm from the catheter tip recorded temperature increases of less than 5.6 °C. This study provides methods for monitoring temperature during in vivo ablation, and demonstrates that functional temperature increases from ablation were restricted to within approximately 5 mm of the catheter. These methods can now be applied to optimize effective ablation parameters, and to inform models of gastric ablation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.