Abstract
Results of neutron transmission Bragg edge spectroscopic experiments performed at the SNAP beamline of the Spallation Neutron Source are presented. A high resolution neutron counting detector with a neutron sensitive microchannel plate and Timepix ASIC readout is capable of energy resolved two dimensional mapping of neutron transmission with spatial accuracy of ∼55 μm, limited by the readout pixel size, and energy resolution limited by the duration of the initial neutron pulse. A two dimensional map of the Fe 110 Bragg edge position was obtained for a bent steel screw sample. Although the neutron pulse duration corresponded to ∼30 mÅ energy resolution for 15.3 m flight path, the accuracy of the Bragg edge position in our measurements was improved by analytical fitting to a few mÅ level. A two dimensional strain map was calculated from measured Bragg edge values with an accuracy of ∼few hundreds μistrain for 300s of data acquisition time.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.