Abstract

In paediatrics, the risks associated with ionising radiation should be a major concern, due to children's higher susceptibility to radiation effects. Measure entrance skin dose (ESD) in chest and pelvis X-ray projections and compare the results with the 'European guidelines on quality criteria for diagnostic radiographer images in paediatrics' in order to optimise radiological practice. ESD values were obtained using an ionisation chamber Diamentor M4 KDK (PTW) in 429 children, who underwent chest X-ray or pelvis X-ray in a Computed Radiography system. In the first phase of the study, data were collected according to protocols used in the department; in a second phase different tube voltage values were used according to patient weight. A third phase was carried out, only for chest X-ray, using the exposure parameters of phase 2, plus activating lateral ionisation chamber. Three paediatric radiologists blindly assessed image quality of chest X-ray, using a validated assessment available in the 'European guidelines on quality criteria for diagnostic radiographer images in paediatrics'. Considering all the patients submitted to chest X-ray, the average ESD was 0.22, 0.16 and 0.08 mGy, for phases 1, 2 and 3, respectively. For pelvis X-ray, the average ESD decreased from 1.18 mGy in phase 1 to 0.78 mGy in phase 2. Dose optimisation was achieved. ESD was reduced 63.6 and 33.9 % in chest and pelvis X-ray, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.