Abstract

Fetal development is essential to the human lifespan. As more and more multifetal gestations have been reported recently, clinical diagnosis using magnetic resonance imaging (MRI), which introduced radiofrequency (RF) exposure, raised public concerns. The present study developed two whole-body pregnant models of 31 and 32 gestational weeks (GWs) with twin fetuses and explored RF exposure by 1.5 and 3.0T MRI. Differences in the relative position of the fetus and changes in fetal weight can cause differences in fetal peak local specific absorption rate averaged over 10g tissue (pSAR10g). Variation of pSAR10g due to different fetal positions can be ~35%. Numerically, twin and singleton fetal pSAR10g results were not significantly different, however twin results exceeded the limit in some cases (e.g. fetuses of 31 GW at 1.5T), which indicated the necessity for further research employing anatomically correct twin-fetal models coming from various GWs and particular sequence to be applied.