Abstract

We have investigated resistively shunted direct-coupled high-Tc dc superconducting quantum interference device (SQUID) magnetometers with four different inductances (50, 100, 150, and 200 pH). The SQUIDs were based on 200 nm thick YBa2Cu3O7−δ films deposited on bicrystal substrates with a 24° misorientation angle, and the shunt resistors were fabricated by masked ion damage. At T=77 K, good quantitative agreement was observed between the measured maximum voltage modulation depth ΔV and calculated values based on the theoretical predictions by Enpuku et al., whereas the white magnetic flux noise SΦ at 10 kHz of all four devices was found to be a factor of 2.3 higher than predicted. The lowest white magnetic field noise of 153 fT/Hz was obtained for the magnetometer with a SQUID inductance of 100 pH with an outer dimension of the pickup loop of just 2 mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.