One hundred ninety-one simultaneous PtcCO2 and PaCO2 values were recorded from 20 neonates using a heated CO2 sensor. Gestational age and birth weight ranged from 30--44 weeks and 1.22--3.4 kg, respectively. An overestimation of PaCO2 was consistently found. Regression analysis revealed a slope of 1.29, an intercept of 18 mm Hg and a correlation coefficient of 0.85 (p less than 0.001). The PtcCO2 sensor has multiple potential uses in the sick neonate but further experience is necessary to provide adequate correction factors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.