Abstract

This study aimed to determine the detection rate of transcranial color-coded sonography (TCCS) of cerebral veins and sinuses and to explore the diagnostic accuracy of TCCS for straight sinus (SS) and transverse sinus (TS) thromboses. The detection rates of cerebral veins and sinuses using TCCS and contrast-enhanced TCCS (CE-TCCS) were analyzed. The diagnostic accuracy of CE-TCCS was evaluated. Median time from symptoms to CE-TCCS was 10 (range, 1–150) d. The detection rate of bilateral basal veins of Rosenthal was 100% by CE-TCCS, followed by right TS (91.89%), SS (88.12%), left TS (74.59%) and vein of Galen (70.27%). Compared with magnetic resonance imaging/magnetic resonance venography, CE-TCCS showed 100% sensitivity and 96.3% specificity for SS thrombosis, 100% and 100% for right TS thrombosis and 100% and 94.4% for left TS thrombosis. In conclusion, CE-TCCS shows high identification rates of cerebral veins and sinuses and a high diagnostic accuracy for SS and TS thrombosis.