Abstract

The paper focuses attention on the development of a control system for a new radiotherapy treatment machine that makes use of a betatron to deliver electron beams over the whole surface of a patient. The betatron moves above a patient at a fixed distance from the patient support system. This paper presents a method to predict the betatron trajectory, in terms of speed and position, according to variation in the patient's geometry and is concerned with a simulation study of a PID control strategy implemented to control the speed of the betatron.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.