Abstract

In this paper, we present a new substitution method based on the unique properties of synchrotron radiation and of the Metrology Light Source (MLS), the dedicated electron storage ring of the PTB. The MLS is used as a light source with a dynamic range of its photon flux of 11 orders of magnitude to bridge the gap in optical power measurement between a cryogenic electrical substitution radiometer and a single photon detector. Two single photon avalanche diodes were calibrated at 651 nm with combined relative uncertainties of 0.17% and 0.16% traceable to a primary standard, a cryogenic electrical substitution radiometer.