Abstract

Radiometric calibrations of fibre-coupled single photon detectors are experiencing growing demand, especially at the telecommunication wavelengths. In this paper, the radiometric calibration of a fibre-coupled superconducting nano-wire single photon detector at the telecom wavelength 1.55 µm by means of well-characterized synchrotron radiation is described. This substitution method is based on the unique properties of synchrotron radiation and the Metrology Light Source, the dedicated electron storage ring of the Physikalisch-Technische Bundesanstalt, and is suitable for fibre-coupled single photon detectors. The Metrology Light Source is used as a light source with a high dynamic range of the radiant power to bridge the radiometric gap occurring in the transition from radiant power measurements and the counting of photons with single photon detectors. Very low uncertainties below 2% have been achieved in the measurement of the detection efficiency of a fibre-coupled superconducting nano-wire single photon detector.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.