Abstract
Induced-fit responsive dopamine (DA) imprinted polymer, poly (melamine-co-chloranil), was used as a suitable coating material for the modification of a hanging mercury drop electrode. The zwitterionic conformation of the imprinted polymer responded differential pulse, cathodic stripping voltammetric current, without any false-positive or false-negative contributions of non-specific sorptions, in aqueous environment of complex matrices. The limit of detection (3 σ) of dopamine was found to be as low as 0.148 ng mL −1, by the proposed sensor that could be considered a sensitive marker of dopamine depletion in Parkinson's disease (PD).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.