Abstract
Traces of zinc and bismuth in small amounts of high purity tin (0.5 ∼ 25 mg) were determined by polarized Zeeman AAS with direct atomization of a solid sample in a electrothermal graphite-cup cuvette. Standard solutions of zinc and bismuth were employed for the calibration curve. Absorption intensities (peak areas of the absorption signals) for zinc and bismuth were not influenced by the residual tin in a graphite-cup cuvette, up to 25 mg for zinc and up to 10 mg for bismuth. The relative standard deviations of analytical results were 9 ∼ 43% for 0.002 ∼ 0.05 ppm of zinc and 5 ∼ 10 % for 0.06 ∼ 1.3 ppm of bismuth. Detection limits were 0.012 ng for zinc and 0.03ng for bismuth. These values correspond to 0.0005 ppm of zinc in 25 mg of tin and 0.03 ppm of bismuth in 10 mg of tin. Time (incl. weighing the sample) required for analyzing one sample was about 5 min.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.