Abstract

Abstract A voltammetric method for the determination of iron at detection limit of 4 μg/l is described, using the catalytic current of the reduction of the Fe(III)-triethanolamine (TEA) complex in the presence of bromate ions. the determination was performed at a mercury hanging drop electrode without preconcentration, using the TEA alkaline solution as a supporting electrolyte and the differential pulse technique. A peak current for the Fe(III)-TEA catalytic reduction was observed at a potential of-1.0 V (Ag/AgCl saturated electrode). the influence of TEA, BrO3 and NaOH concentrations on the peak height was studied. It was found that a 100-fold excess of Mn, a 50-fold excess of Cr(VI) and Zn did not interfere in the determination. This method was applied to the determination of iron in water, snow and waste water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.