Abstract

A combined procedure enabling simultaneous multielement analysis of trace impurities in high-purity iron is presented. After removal of the iron matrix by solvent extraction with methyl isobutyl ketone, the trace elements Ti, V, Cr, Mn, Ni, Cu, Pb and Bi are determined by means of total reflection X-ray fluorescence analysis. Detection limits are found in the range of 100 ng/g. The reliability of the method is verified by the analysis of commercial high-purity iron and by the comparison of analytical data obtained by ICP-AES.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.