Abstract

In the present study, we investigated the possibilities and drawbacks of hollow fiber liquid phase microextraction (HF-LPME) combined with total reflection X-ray fluorescence (TXRF) spectrometry for the determination of low amounts of inorganic arsenic (As) species in water samples. The obtained results showed that a three-phase HF-LPME system was more suitable to be used in combination with TXRF than the two phase configuration, since lower detection limit and better precision for As determination can be attained. Relevant experimental parameters affecting As extraction (i.e. types of extractant, organic solvent, agitation speed, pH and extraction time) and TXRF analysis (deposition volume and drying mode) were systematically evaluated. It was found that As(III) was more efficiently extracted at pH 13, whereas, optimum pH for As(V) extraction was at pH 8.5. Limits of detection (LOD) achieved using the best analytical conditions meet the requirements of current legislation and allow the determination of inorganic As(V) and As(III) in water. The proposed method was also applied to different spiked environmental water samples for the preconcentration and subsequent determination of trace inorganic As species.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call