Abstract
The application of Bond-Elut Certify solid-phase extraction columns to the systematic toxicological analysis of whole blood was evaluated. The reproducibility of the extraction was tested with thirteen drugs varying in physico-chemical properties. Analysis was performed with capillary gas chromatography with nitrogen-selective detection. The recoveries were reproducible, as long as other limiting factors, e.g., chromatographic behaviour or volatility, do not play a significant role. The effect of limiting chromatographic behaviour was studied in more detail with the more sensitive mass spectrometry with selected-ion monitoring after converting the extracted morphine into its ditrimethylsilyl derivative.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.