Abstract
We report the instrumentation of a new positron diffractometer that is connected to the high-intensity positron beam at the neutron induced positron source Munich. Crucial elements for the adaption of the positron beam are presented, which include the magnetic field termination, the optional transmission-type remoderator for brightness enhancement, and the electrostatic system for acceleration and beam optics. The positron trajectories of the remoderated and the twofold remoderated beam have been simulated to optimize the system, i.e., to obtain a coherent beam of small diameter. Within a first beamtime, we tuned the system and characterized the direct beam. For the twofold remoderated beam of 10keV energy, we experimentally observe a beam diameter of d < 1.3mm, which agrees well with the simulation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.