Abstract

An isocratic high-performance liquid chromatography with electrochemical detection (HPLC–ED) method for the determination of total plasma homocysteine [H(e)] has been developed. The electrochemical detection is performed using a glassy-carbon electrode that is not specific for thiol groups. We have tried to solve the problem of specificity focusing our work on chromatographic resolution and have obtained good results without coelution of other thiol compounds or any substances mentioned as common interferences for carbon electrode methods: uric acid, ascorbic acid and salicylates. Thirty samples a day can be assayed for total homocysteine with a lower limit of detection of 2 pmol, and a limit of quantification of 1.0 μmol/l, with a coefficient of variation (C.V.) <20%. For a concentration of total plasma homocysteine of 9.36 μmol/l, the intra- and inter-assay C.V.s were of 3.86% and 5.55% respectively. The analytical recovery achieved in the preparation of the samples ranged from 85.0% to 98.3% and the electrochemical response was linear up to 100 μmol/l.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.