Abstract

Abstract The BactoScan 8000 (Foss Electric) automatically performs a total count of bacteria in raw milk. It was designed for fast, repeatable counting. The apparatus counts the individual bacteria (IBC) instead of colony forming units (CFU). IBC provides more information on the state of milk hygiene than the CFU count. The repeatability, sr, was 0.033 log (IBC/μL), which is equivalent to a coefficient of variation of 7.9% (IBC/μL). The carryover, which is the contamination one sample imposes on the next, was 0.19%. The BactoScan 8000 was calibrated against the reference method of counting colonies on a Petri dish. The correlation coefficient for 429 milk samples was r= 0.88, and the standard deviation of the residuals, syx, was 0.259. The detection limit was 15 CFU/μL. We recommend establishing a system of payment to farmers based on the bacteriological quality of their milk as determined by the BactoScan 8000. Because results are obtained in IBC/μL, analytical differences resulting from conversion to CFU/|iL are eliminated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.