Abstract

Today almost a quarter of all neutron scattering experiments performed at neutron scattering facilities require sample temperatures below 2 K. However, a global shortage of helium gas can seriously jeopardise low temperature experimental programmes at neutron scattering laboratories. Luckily the progress in cryo-cooler technology offers a new generation of cryogenic systems with significantly reduced consumption and in some cases nearly a complete elimination of cryogens. Here we discuss design and test results of a new cryogen free top-loading cryostat developed through an ISIS and Oxford Instruments collaborative project. The cryostat provides neutron scattering sample environment in the temperature range 1.4–300 K. High cooling power (0.23 W at temperature less than 2 K) achieved at the cryostat’s variable temperature insert heat exchanger allows operation of a standard dilution refrigerator insert in a continuous regime. From a user perspective, the system offers operating parameters very similar to those of an Orange cryostat but without the complication of cryogens.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.