Abstract
The Silicon Photomultiplier (SiPM) is intrinsically a very fast device, its single photoelectron timing resolution is about 100 ps FWHM. Therefore real timing properties of the system scintillator+SiPM is determined mostly by timing properties of the scintillator+light collection system. We present the experimental results for timing properies of SiPM+scintillator (or Cherenkov radiator) for two cases: (1) timing resolution in a few GeV electron beam with fast plastic scintillator or Cherenkov radiator+SiPMs with a size of 1×1 mm 2 and 3×3 mm 2 and (2)time-of-flight resolution for detection of 511 KeV photons by two LSO crystals+SiPM (3×3 mm 2), as a possible application for PET.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.