Abstract

This work reports on Silicon Photomultipliers (SiPM) timing resolution measurements performed at the picosecond level at Laboratory of Linear Accelerator (LAL), In2p3- CNRS.The dependence of Single Photoelectron Timing Resolution (SPTR) with the applied voltage, wavelength of the light and the temperature was measured for detectors from Hamamatsu Photonics, AdvanSiD and Sensl with an active area of 1 and 9mm2.The SPTR improves with the bias voltage increase. No significant variation of SPTR was observed with the temperature change. We also observed a weak variation of it as a function of the wavelength of the light. The best SPTR measured was about 120ps (FWHM).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call