Abstract
This paper presents a conceptually new design of the 57Fe Time Differential Mössbauer Spectrometer (TDMS) with the gamma-photon detector optimized for registration of a radiation emitted in a maximum solid angle. A high detection efficiency of 80% in 4π region was achieved for 122.06keV photons emitted from 57Co source. Detector parameters have been optimized for the use in the Time Differential Mössbauer Spectroscopy where the high time resolution in range of 176–200ns is highly required. Technical concept of the TDMS is based on the virtual instrumentation technique and uses fast digital oscilloscope. Performance and detector utilization have been clarified by decreasing the Mössbauer spectral line-width of K2MgFe(CN)6 reference sample from 0.33mm/s (integral mode) to 0.23mm/s (time differential mode). This report also describes characterization and utilization of the detector together with additional electronic blocks and two-channel fast data-acquisition system construction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.