Abstract

Temperature sensor calibration in the majority of measurement laboratories is performed on the ground of measurement procedures. The procedures specify the timespan for the temperature of the tested sensor to stabilize. In practice, the fact of time constant increase above expectations can be observed. This results from the method of operation, time of use, and level of contamination of individual sensors. The paper proposes a method to optimize the calibration process of resistance temperature sensors based on the measurement of the time constant and, on this basis, to determine the sensor stabilization time during calibration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.