Abstract

Purpose : Locking screw deformation is common in nailing of comminuted tibia fractures. The aim is to compare five different proximal locking screws bending resistance in tibia nailing system. Methods : 50 screws were divided into five groups. A stainless steel tube which has a 34 mm internal diameter was prepared representing the proximal tibia. A 3-point bending tests were performed on locking screws for determining their yield points with 3 different dimensions (4.5 mm, 5 mm and 5,5 mm) and with 2 different screw thread (low threaded and unthreaded). Results : The mean yield point values of smooth 4.5 mm and threaded 5 mm low locking screws were statistically significant less than that of smooth 5 mm, low threaded 5.5 mm and smooth 5.5 mm,(P=0.000). Conclusions : To avoid proximal locking screw deformation, use of smooth 4,5 mm and low threaded 5 mm locking screws should be avoided in nailing of comminuted tibia fractures of unreliable persons. Smooth 5 mm, low threaded 5.5 mm and smooth screws 5.5 mm may be used safely.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.