Abstract

A new nephelometer suitable for aircraft measurements of aerosol scattering extinction coefficient (σsp) has been constructed and operated under field conditions. This instrument is vacuum tight for operation in a pressurized aircraft cabin and is capable of measuring background tropospheric σsp at an averaging time of 1 min. For example, in a typical atmospheric profile the instrument can measure values of about 10−4 m−1 with a time resolution of 2 s in a polluted region, and about 10−7 m−1 with a time resolution of 1 min in a clean region. This sensitivity is made possible by: (1) subtracting in real time the air Rayleigh scattering from the total scattering signal by continuously measuring pressure and temperature in the sampling volume of the instrument; (2) correcting for the dark count and sensitivity of the photomultipliers using a rotating shutter; and (3) using a beam splitter arrangement to allow simultaneous detection by three photomultipliers. A laboratory measurement of instrument noise suggests a 550-nm noise level of about 5 × 10−8 m−1 at an averaging time of 1 min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.