Abstract
We adapted three bichromatic spectrophotometric methods for determining total bilirubin in serum, for use with the Technicon RA-1000 analyzer. The borate buffer (BOR) of Hertz et al. (Scand J Clin Invest 1974;33:215-30), the caffeine buffer (CAF) of Vink et al. (Clin Chem 1988;34:67-70), and the combined borate-caffeine buffer (B-C) of Franzini and Cattozzo (Clin Chem 1987;33:597-9) were compared. All methods required only 10 microL of serum, were precise (between-batch CVs less than 4.2%, analyte range 64-310 mumol/L), linear to 1000 mumol/L, and insensitive to interference from hemoglobin to 5 g/L. Lipemia, carotene, and methemalbumin interfered positively with each method but insignificantly unless in supranormal concentrations. Only the BOR method was sensitive to protein matrix effect. Neonates' results obtained with all three methods compared well with a Jendrassik-Gróf-based technique. However, samples from adults with cholestasis were overestimated, particularly by the CAF method, but the BOR and B-C methods would be suitable for "stat" bilirubin analysis in these samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.