Abstract

A polymer hydrogel foam is proposed as a potential three dimensional experimental dosimeter for radiation treatment verification in low-density tissue such as the lung. A gel foam is created by beating a radiation sensitive polymer gel mixture in an anoxic atmosphere. The mass density of the gel foam is in the order of 0.25-0.35 kg/dm3. Both nuclear magnetic resonance (NMR) spin-spin relaxation rate (R2) and magnetization transfer ratio (MTR) have been used to map the dose distribution from the gel dosimeter. It is found that MTR has significant advantages compared to R2 for mapping the dose distribution in the polymer gel foam dosimeters. The magnetization transfer ratio is found to be less dependent on the density and microstructure of the gel foam dosimeter while spin-spin relaxation dispersion has been observed making the spin-spin relaxation rate dependent on the interecho time interval. Optical microscopy reveals a microstructure that shows great similarity with human lung tissue. It is also shown how NMR hydrogen proton density measurements can be used to map the density distributions in gel dosimeters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.