Abstract

A novel composition of normoxic polymer gel dosimeters based on radiation-induced polymerization of N-(Hydroxymethyl)acrylamide (NHMA) is introduced in this study for 3D dosimetry for Quality Assurance (QA) in radiation therapy. Dosimeters were irradiated by 6, 10 and 18MV photon beams of a medical linear accelerator at various dose rates to doses of up to 20Gy. The dose response of polymer gel dosimeters was studied using nuclear magnetic resonance (NMR) spin-spin relaxation rate (R2) of hydrogen protons within the water molecule. Also, we measured gel response using absorption spectroscopy and found that this novel gel can be successfully utilized for both MRI- and OCT- (Optical Computed Tomography) based 3D dosimetry. We investigated dosimetric properties of six different compositions of the new NHMA-based gel in terms of dose rate, radiation beam quality and stability of dose-dependent polymerization after irradiation. We found no significant effects of these parameters on the novel gel dosimeter performance in both relaxation rate and absorbance measurements.