Abstract

We report a new implementation of a recently developed 3D momentum imaging technique [Lee etal. J. Chem. Phys. 141, 221101 (2014)]. The previously employed high-speed digitizer in the setup is replaced by a portable USB3 oscilloscope. A new triggering scheme was developed to suppress trigger jitters and to synchronize the signals from a camera and the oscilloscope. The performance of the setup was characterized in the study of laser desorption/ionization of 2,5-dihydroxybenzoic acid on a velocity map imaging apparatus. A ∼60 picosecond time resolution in measuring time-of-flight is achieved with a count rate of ∼1kHz, which is comparable to the system using high-speed digitizers. The new setup affords great portability and wider accessibility to the high-performing 3D momentum imaging technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.