Abstract

Background: Clinicians often rely on measurement of carboxyhemoglobin (COHb) to confirm or rule out a diagnosis of carbon monoxide (CO) poisoning. Methods: We report two cases of false negative COHb in patients with CO poisoning and one case of false positive COHb in a patient without CO poisoning. Results: In the first case, a 20-year-old male developed headache, confusion, and near-syncope while operating a gasoline-powered pressure washer in an enclosed space. In the emergency department (ED), his COHb was 1.8%, but this level was disregarded, and he was referred for hyperbaric oxygen. His COHb just before hyperbaric oxygen was 4.1%, and later analysis of his blood collected at ED arrival revealed a COHb of 20.1%. The referral ED blood gas machine calibration and controls were within specification. In the second case, a 45-year-old male presented with several others to the ED with symptoms of CO poisoning after exposure at a conference. All others had elevated COHb levels, but his COHb was 2%. He was discharged but returned shortly with continued symptoms and requested his COHb be repeated. The repeat COHb was 17% (84 minutes after the first). After three hours of oxygen, his COHb was 7%. In the final case, an 83-year-old non-smoking male presented to an ED with breathlessness and tachypnea and was diagnosed with COVID-19 pneumonia. His COHb was 7.1%, but he reported living in an all-electric home. Another adult who lived with him and rode with him to the ED was asymptomatic and had a COHb of 3%. Later, COHb of 1.9% was measured from blood collected at ED arrival, and gas chromatography/mass spectrometry confirmed this result (2%). Conclusions: COHb levels are not always accurate. Clinicians should use clinical judgment to manage their patients, including rejecting laboratory values that do not fit the clinical situation.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call