In this work, we describe a novel method to improve the detection limits of the non-classical polymeric membrane ion-selective electrodes (ISEs) which are conditioned with highly discriminated ions instead of primary ions. It is based on a thin-layer ISE membrane with a thickness of 5 μm, which is coated on ordered mesoporous carbon used as solid contact. The diffusion of the primary ion from the surface of the sensing membrane to the bulk of the membrane could be avoided by the proposed thin membrane configuration. Since the detection sensitivity of the non-classical ISEs depends on the accumulation of the primary ion in the interfacial layer of the sensing membrane, a lower detection limit can be obtained. By using the copper ion as a model, the present potentiometric sensor shows a significantly improved detection sensitivity compared to the conventional ISE with a membrane thickness of ca. 200 μm. Low detection limits of 0.29 and 0.53 nM can be obtained in 0.01 and 0.5 M NaCl, respectively. In addition, the proposed sensor exhibits an excellent reversibility by using a neutral proton-selective ionophore incorporated in the thin membrane.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.