Abstract
Two types of thermoluminescent dosimeters (TLDs): Harshaw LiF:Mg,Ti (TLD-100) and CaF 2 :Tm (TLD-300) were investigated for their glow curve responses to separate, mixed field, and simultaneous X-ray and proton irradiations. All TLDs were exposed to X-rays produced by a 30 keV, 0.2 mA electron beam, X-ray tube source. TLD-100 detectors were exposed to 2.5 MeV protons and the TLD-300 detectors were exposed to 4 MeV protons. The following five separate irradiation schemes were used: only X-ray, only proton, first X-ray then proton, first proton then X-ray, and simultaneous X-ray and proton beam exposures. Computerized glow curve deconvolution based on first order TL kinetics was used to analyze the resulting glow curves for any differences in shape and TL efficiency due to order of exposure. We found no significant differences in glow curve shape or magnitude attributable to radiation exposure order.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.