Abstract

Two types of thermoluminescence dosemeters (TLDs), the Harshaw LiF:Mg,Ti (TLD-100) and CaF(2):Tm (TLD-300) were investigated for their glow curve response to separate photon and proton irradiations. The TLDs were exposed to gamma irradiation from a (137)Cs source and proton irradiation using a positive ion accelerator. The glow curve peak structure for each individual TLD exposure was deconvolved to obtain peak height, width, and position. Simulated mixed-field glow curves were obtained by superposition of the experimentally obtained single field exposures. Feature vectors were composed of two kinds of features: those from deconvolution and those taken in the neighbourhood of several glow curve peaks. The inner product of the feature vectors was used to discriminate among the pure photon, pure proton and simulated mixed-field irradiations. In the pure cases, identification of radiation types is both straightforward and effective. Mixed-field discrimination did not succeed using deconvolution features, but the peak-neighbourhood features proved to discriminate reliably.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.