Abstract

Dosimetry for intraoperative radiotherapy (IORT) after wide local excision for breast cancer using a 50 kV X-ray needle (Intrabeam) was performed in vivo using thermoluminescence dosimetry. Eight LiF:Mg,Ti chips were placed on the skin around the incision site after wide local excision while the tumour bed was irradiated to a prescribed dose of 5 Gy 10 mm from the applicator surface. The maximum and mean measured skin dose for 57 patients ranged from 0.64 to 7.1 Gy and 0.56 to 4.78 Gy, respectively, reflecting different tissue thicknesses overlying the applicator. The average maximum dose of 2.93+/-1.46 Gy was below the threshold for severe radiation skin toxicity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.