Abstract

Thermodynamic temperatures during the melt and the freeze of Co–C, Pd–C, Pt–C and Ru–C metal–carbon fixed-point cells manufactured by LNE–INM/CNAM, NMIJ and NPL were determined by absolutely calibrated filter radiometers traceable to the PTB cryogenic radiometer and a radiance comparison method using an IKE LP3 radiation thermometer. The measurement uncertainties were below 400 mK at temperatures up to 2250 K. The results are in agreement within the combined uncertainties with a study on relative temperature differences of the same set of fixed-point cells. For the fixed-point cells manufactured by NPL the results are compared with a previous thermodynamic temperature measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.