Abstract

Five iron–carbon (Fe–C) eutectic fixed-point cells have been constructed between NPL and LNE-Cnam to investigate the robustness and to measure the agreement of their melting temperatures. Each cell was constructed with a different selection of materials sourced by NPL and LNE-Cnam. The measured emfs at the Fe–C fixed-point temperature (∼1153 °C), compared between cells, agree within around 1.98 µV (∼90 mK), where the most important contribution to the uncertainty of each measurement is the inhomogeneity associated with the measuring Pt/Pd thermocouple. This demonstrates that these cells are suitable for use as secondary fixed-point cells in contact thermometry but the robustness of the presented cells is not found to be sufficient for maintaining their integrity during repeated cycling procedures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.