Abstract
We report on a new solid state dosimeter based on chemical vapor deposition (CVD) single crystal diamond fabricated at Roma “Tor Vergata” University laboratories. The dosimeter has been specifically designed for direct neutron dose measurements in boron neutron capture therapy (BNCT). The response to thermal neutrons of the proposed diamond dosimeter is directly due to 10B and, therefore, the dosimeter response is directly proportional to the boron absorbed doses in BNCT. Two single crystal diamond detectors are fabricated in a p-type/intrinsic/metal configuration and are sandwiched together with a boron containing layer in between the metallic contacts (see Fig.1). Neutron irradiations were performed at the Frascati Neutron Generator (FNG) using the 2.5 MeV neutrons produced through the D(d,n) 3He fusion reaction. Thermal neutrons were then produced by slowing down the 2.5 MeV neutrons using a cylindrical polymethylmethacrylate (PMMA) moderator. The diamond dosimeter was placed in the center of the moderator. The products of 10B(n,α)Li nuclear reaction were collected simultaneously giving rise to a single peak. Stable performance, high reproducibility, high efficiency and good linearity were observed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.