Abstract

The use of infrared thermography to evaluate the perfusion of tissue flaps have been studied. This study aims to evaluate the utility of thermal imaging for flap monitoring with a low-cost, readily available smartphone imaging device. Adult subjects who underwent head and neck reconstruction using a microvascular free flap with a cutaneous paddle were recruited. Thermal images were taken of the free flap before, during and after anastomosis. Thermal images were analyzed by measuring the average flap temperature minus the average surrounding tissue temperature (dT). Twenty-one patients were enrolled. The mean dT for flaps intraoperatively prior to anastomosis was -11.47 °F. For 20 patients, dT averaged between -0.30 to 0.12 °F. One flap was inadequately perfused and dT was found to be -4.35 °F. Low cost, mobile smartphone devices such as the thermal camera may provide an objective method of monitoring microvascular free flaps. 2 Prospective Cohort Study.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.