Abstract

3MeV H− beam from the Radio Frequency Quadrupole (RFQ) will be accelerated to 80MeV in the CSNS (China Spallation Neutron Source) linear accelerator (Linac). The wire scanner is used to measure the transverse beam profile and the emittance, and the carbon or tungsten wire is considered to use. Thermal analysis of the wire scanners in the Linac is presented in this paper. The maximum temperature (Tm) of the wire decreases as the beam energy increases, and we also calculate the influence of all possible parameters on Tm. Tm of carbon wire is significantly lower than tungsten wire if both the beam parameters and wire geometric parameters are set to the same, which can be attributed to its higher heat capacity and radiant emissivity. In addition, we present the results of sublimation rate of the wire, which show that tungsten wire has a much lower evaporation rate than carbon wire in the same temperature, which can be attributed to the different vapor pressures of the two materials. To limit the thermionic emission, the maximum beam frequency approximately has an exponential relationship with beam rms size at a certain beam pulse width.