Abstract

The high flux density of synchrotron white beam offers several advantages in X-ray imaging such as higher resolution and signal-to-noise ratio in 3D/4D micro-tomography, higher frame rate in real-time imaging of transient phenomena, and higher penetration in thick and dense materials especially at higher energies. However, these advantages come with additional challenges to beamline optics, camera and sample due to increased heat load and radiation damage, and to personal safety due to higher radiation dose and ozone gas hazards. In this work, a white beam imaging facility at imaging beamline BL-4, Indus-2, has been developed, while taking care of various instrumental and personal safety challenges. The facility has been tested to achieve 1.5 µm spatial resolution, increased penetration depth up to 900 µm in steel, and high temporal resolutions of ∼10 ms (region of interest 2048 × 2048 pixels) and 70 µs (256 × 2048 pixels). The facility is being used successfully for X-ray imaging, non-destructive testing and dosimetry experiments.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.