Abstract
Small bowel carcinoid (SBC) primary tumors can be multifocal in 40%-55% of patients and challenging to detect. Near infrared autofluorescence (NIRAF) is used for detection of parathyroid glands. It is unknown if this technology can be used to identify SBCs and how it would compare with current imaging modalities. This was a prospective institutional review board-approved pilot study of three patients undergoing resection of SBCs. NIRAF was used to image SBCs and mesenteric lymph nodes intraoperatively and at back table. Findings were compared with preoperative imaging, surgical exploration and pathology. Statistics were performed using Mann-Whitney U test. Eleven SBCs and 12 mesenteric lymph nodes were analyzed. All SBCs had a brighter focal autofluorescence (AF) signal compared to background. Normalized pixel intensity of SBCs was 2.2 (0.7) and normal small bowel 1.4 (0.6) (p < 0.0001). NIRAF was less accurate in detecting occult lymph node metastasis, but was superior to DOTATATE PET for detecting SBCs in two of three patients. This preliminary report suggests that SBCs exhibit distinctly bright AF signals on NIRAF to create a contrast distinction from normal small bowel. This technology may have a utility as an adjunctive tool for intraoperative detection of occult SBCs.