Abstract

BackgroundExperience treating proximal humerus fracture sequelae with reverse total shoulder arthroplasty is limited. We report our results. PatientsForty-four patients with sequelae of a proximal humeral fracture were treated with a reverse total shoulder prosthesis. There were 26 women and 18 men, with a mean age of 77years (range, 74–84years). The mean follow-up after reverse arthroplasty was 48months (range, 40–84months). ResultsThe mean Constant score increased from 28 preoperatively to 58 postoperatively (p < 0.0001). The average anterior elevation increased from 40° to 100° (p < 0.0001), abduction from 41° to 95° (p < 0.0001), external rotation from 15° to 35° (p < 0.0001) and internal rotation from 25° to 60° (p < 0.0001). The average subjective shoulder score increased from 13% preoperatively to 56% postoperatively (p < 0.0001). All but six patients would undergo the same procedure again if faced with the same problem. Twenty-four patients were very satisfied, 14 satisfied and 6 unhappy with the operation. Six prosthetic dislocations occurred (13.6%). Two of them were successfully treated by adding an extension to the humeral neck component to increase the offset and tension. In the other four dislocations this procedure failed, and the prosthesis was revised and converted to a hemiarthroplasty. There was one case of glenoid component loosening that was converted to a hemiarthroplasty. ConclusionsThe reverse total shoulder arthroplasty improves function and motion in patients with proximal humeral fracture sequelae. However, the dislocation rate is high.