Abstract

The use of DSC curves is proposed as an alternative method to determine the degree of N-acetylation (DA) in chitin/chitosan samples, based in both peak area and height of the decomposition signal. Samples with DA from 74 to 16% were prepared from a chitin commercial sample and the DA was determined by 1H NMR, 13C CP/MAS NMR and IR spectra. The effect of water content, heating rate, sample mass and gas flow on the DSC peaks were evaluated and optimized. Using optimized conditions a linear relationship between peak area and height with the DA could be achieved with linear correlation coefficients of −0.998 and −0.999 (n=7), respectively. The calibration graphs were used to determine the DA of a commercial chitosan sample with relative errors ranging from 2 to 3% for both peak area and peak height, when compared with the DA determined by 1H NMR method.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call